Sale!

CM2201

¥17,541.00

Parameters and specifications

Detection principle: Non-spectral infrared technology (NDIR)
Data output: CO2 concentration (mmHg), end-expiratory CO2, exhaled CO2, respiratory rate
Detection range: 0-150mmHg
Resolution: 0.1mmHg
Detection accuracy:
0- 40mmHg: ±2mmHg
41-70 MMHG: True value ±5%

Category:

Description

Parameters and specifications

Detection principle: Non-spectral infrared technology (NDIR)
Data output: CO2 concentration (mmHg), end-expiratory CO2, exhaled CO2, respiratory rate
Detection range: 0-150mmHg
Resolution: 0.1mmHg
Detection accuracy:
0- 40mmHg: ±2mmHg
41-70 MMHG: True value ±5%
71-100mmHg: True value ±8%
100-150mmHg: True value ±10%
Respiratory rate range: 0-150BPM
Respiratory rate accuracy: ±1 breath

Stability:

Short-term drift: not more than 0.8mmHg(4H)
Long-term drift: Accuracy is maintained within 120H
Input voltage: 5.00V (± 5%)
Average current: <120mA
Peak current: <250mA
Initialization time: Display within 15s, full precision within 2min (25℃)
Size and weight
Size: 52 mm * 23 mm * 36 mm

series

The CM2201 may not be a standalone series, but rather a specific model of Sifang’s end-expiratory ETCO2 sensor.
trait
Accurate and reliable reading: non-spectral infrared detection principle based on independent intellectual property rights design.
Good waveform: Excellent performance in medical applications.
Compact appearance: easy to install and carry.
Flexible installation: applicable to different medical devices and application scenarios.
Compliance standards: Meet the requirements of EN IEC 55014-1:21st, EN IEC 61000-3-2:19th and other standards.